EasyManuals Logo

Waters TQ Detector User Manual

Default Icon
235 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
5-36 Maintenance Procedures
11. Use the needle-nose pliers to grasp the PEEK terminal block and
partially lift it out of the ion block.
12. Holding the PEEK terminal block gently, use the needle-nose pliers to
gently grasp the heat-shrink tubing on the heater cartridge assembly
and slide it and the PEEK terminal block out of the ion block.
Caution: To avoid damaging the heater cartridge assembly
wires, do not bend or twist them either side of the heater
cartridge assembly heat-shrink tubing when removing the
assembly from the ion block.
TP02618
Heat-shrink tubing
Heater cartridge assembly
PEEK terminal block

Table of Contents

Other manuals for Waters TQ Detector

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters TQ Detector and is the answer not in the manual?

Waters TQ Detector Specifications

General IconGeneral
BrandWaters
ModelTQ Detector
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals