5-36 Maintenance Procedures
11. Use the needle-nose pliers to grasp the PEEK terminal block and
partially lift it out of the ion block.
12. Holding the PEEK terminal block gently, use the needle-nose pliers to
gently grasp the heat-shrink tubing on the heater cartridge assembly
and slide it and the PEEK terminal block out of the ion block.
Caution: To avoid damaging the heater cartridge assembly
wires, do not bend or twist them either side of the heater
cartridge assembly heat-shrink tubing when removing the
assembly from the ion block.
TP02618
Heat-shrink tubing
Heater cartridge assembly
PEEK terminal block