Cleaning the ion block, isolation valve, and extraction cone 5-37
13. Use the O-ring removal kit to carefully remove the cover seal from the
ion block (see also “Removing O-rings and seals” on page 5-10).
14. Dispose of the cover seal in accordance with local environmental
regulations.
Warning: The cover seal can be contaminated with
biohazardous and/or toxic materials. Ensure that it is
correctly disposed of according to local environmental
regulations.
TP02619
Cover seal