Replacing the source assembly seals 5-81
8. Use the 5-mm Allen wrench to loosen the 3 captive source enclosure
securing screws.
9. Remove the source enclosure from the instrument.
Caution: Do not apply any downward force to the source
enclosure door when removing the source enclosure from the
instrument’s housing.
RP00027RP00027
Source enclosure
securing screw
Source enclosure
securing screws