Cleaning the sample cone and gas cone 5-21
To remove the cone gas assembly from the source
1. Close the source isolation valve (see “Operating the source isolation
valve” on page 5-8).
2. Disconnect the PTFE tube from the cone gas assembly connection tube.
Warning: The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning: To avoid electric shock, ensure that the instrument is in
Standby mode before commencing this procedure.
Warning: To avoid puncture wounds, take great care while working
with the source enclosure door open if one or both of these conditions
apply:
An ESI probe is fitted (the probe tip is sharp).
A corona pin is fitted (the pin tip is sharp).
Warning: The source can be hot. To avoid burn injuries, take great care
while working with the instrument’s access door open.
Caution: Do not apply any downward force to the source enclosure door
while the door is open.
RP00006
PTFE tube
Cone gas assembly
Connection tube

Table of Contents

Other manuals for Waters TQ Detector

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Waters TQ Detector and is the answer not in the manual?

Waters TQ Detector Specifications

General IconGeneral
BrandWaters
ModelTQ Detector
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals