19. INFRARED CAMERA BEHAVIOR
19.3 Camera Calibraon
The WIRIS Pro undergoes external calibraon as a nal step of the manufacturing process and the user doesn’t
need to perform addional calibraon other than the automac NUC.
The stated accuracy of a calibrated camera is ±2% of measured value or ±2°C (±3.6°F) absolute (whichever is
higher) on the measurement range of 0 to 500°C (32°F to 932°F) and with camera already warmed up with
ambient temperature in the range from 0°C to 40°C (32°F to 104°F).
Note: For maximal accuracy to be maintained, it’s recommended for the device to be sent to the manu‐
facturer for a calibraon once a year, otherwise the stated accuracy cannot be guaranteed.
Warning: as with all measurement devices, the volality of the device’s storage environment in terms
of temperature swings, light radiaon, electromagnec interference and vibraons can have detrimental
eects on quality of the device’s calibraon and its longevity. Aempt to store the device with as few
disturbances of any sort as possible for maintaining maximal accuracy.