6-17
IM DL850-03EN
Specifications
1
2
3
4
5
6
App
Index
Item Specifications
Cable conditions (input module)
Isolated probe 700929 (for the 701250, 701251, 701255, 701275, and 720210)
Wrap the cable once around a ferrite core
6
on the side of the cable closest to the DL850/
DL850V.
Twisted pair cable for the 701261, 701262, and 701265
Use a cable that is 3 m or less in length, and wrap the cable once around a ferrite core
6
on
the side of the cable closest to the DL850/DL850V.
Twisted pair cable for the 720220 and 720221(701953)
Use a cable that is 3 m or less in length, and wrap the cable once around a ferrite core
8
on
the side of the cable closest to the DL850/DL850V.
Measurement lead 758933 (for the 701267)
Bundle the H and L measurement leads. Attach a ferrite core
6
to the DL850/DL850V side of
the bundled leads.
D-sub 9-pin shielded cable (for the 720240 and 720241)
Use a cable that is 3 m or less in length, and attach a ferrite core
6
to the side of the cable
closest to the DL850/DL850V.
Bridgeheads 701955 and 701956 (for the 701270)
Use a cable that is 5 m or less in length, and attach a ferrite core
6
to the side of the cable
closest to the DL850/DL850V.
Bridgeheads 701957 and 701958 (for the 701271)
Use a cable that is 5 m or less in length, and attach a ferrite core
6
to the side of the cable
closest to the DL850/DL850V.
Logic probe input cables 700986, 700987, 702911, and 702912 (for the 720230)
Attach a ferrite core
6
to the side of the cable closest to the DL850/DL850V.
Immunity Compliant standards
EN61326-1 Table 2 (for use in industrial locations), EN61326-2-1
Approved (DL850/DL850V, 701250, 701251, 701255, 701267, 701261, 701262, 701265,
701270, 701271, 701275, 701280, 720210, 720220, 720221, 720230, 720240, 720241,
700986, 700987, 701953, 701955, 701956, 701957, 701958, 702911, 702912)
Influence in the immunity environment (criteria A)
Noise increase
701250: ≤±20 mV (1:1 input, 5 mV/div conversion)
701251: ≤±3 mV (1:1 input, 1 mV/div conversion)
701255: ≤±25 mV (1:1 input, 5 mV/div conversion)
701267: ≤±30 mV (1:1 input, 20 mV/div conversion)
701261: ≤±3 mV (5 mV/div range conversion)
701262: ≤±3 mV (5 mV/div range conversion)
701265: ≤±0.05 mV (0.1 mV/div)
701270: ≤±100 μSTR (±500 μV range, gauge factor = 2)
701271: ≤±100 μSTR (±500 μV range, gauge factor = 2)
701275: ≤±6 mV (1:1 input, 5 mV/div range conversion)
701280: ≤±0.01 Hz (frequency, 0.1 Hz/div conversion)
720210: ≤±50 mV (1:1 input, 10 mV/div range conversion)
720220: ≤±20 mV (0.2 V/div range conversion)
720221: ≤±2 mV (1 mV/div range; data update period = 100 ms when you are using
the 701953)
720230: No bit errors
720240: No erros
720241: No erros
Test conditions
701250: 10 MS/s, envelope mode, 50 mV/div, no input filter, with the tip of the probe
(700929 (10:1)) shorted
701251: 1 MS/s, envelope mode, 10 mV/div, no input filter, with the tip of the probe
(700929 (10:1)) shorted
701255: 10 MS/s, envelope mode, 50 mV/div, no input filter, with the tip of the probe
(701940 (10:1)) shorted
701267: 100 kS/s, envelope mode, 0.2 V/div, no input filter, with the end of the cable
shorted
701261: 100 kS/s, envelope mode, 5 mV/div, no input filter, with the end of the cable
shorted
701262: 100 kS/s, envelope mode, 5 mV/div, no input filter, with the end of the cable
shorted
701265: 500 S/s, envelope mode, 0.1 mV/div, no input filter, with the end of the cable
shorted