6-38
IM DL850-03EN
Item Specifications
Input voltage range (±FS) When using 1:1 probe attenuation: ±1 V, ±2 V, ±5 V, ±10 V, ±20 V, ±50 V (±FS)
Input impedance 1 MΩ ± 1 approx. 35 pF
Pull-up function: 4.7 kΩ, approx. 5 V (pull-up can be turned ON only when the
input is set to Pull-Up 5 V)
Input coupling settings AC, DC
Probe attenuation setting 10:1, 1:1
Minimum voltage width for pulse
detection
200 mV
P-P
Bandwidth limit Select from Full, 100 kHz, 10 kHz, 1 kHz, and 100 Hz
Cutoff characteristics: −12 dB/OCT (Typical
9
)
Threshold Set within the FS of the voltage range. Set in units of 1% of the FS.
Hysteresis Select ±1%, ±2.5%, or ±5% of the FS of the voltage range
Preset function Logic (5 V/3 V/12 V/24 V), electromagnetic pickup, zero crossing, pull-up, AC100 V, AC200 V,
and user-defined
Slope selection Select rising or falling
Lower −3 dB point when AC
coupled
0.5 Hz or less (0.05 Hz or less when using the 700929, 0.005 Hz or less when using the
701947) (Typical
9
)
Chatter elimination function OFF or 1 to 1000 ms (1 ms resolution)
Eliminates the chatter that occurs such when the contact input is turned ON/OFF.
Can discard the signal changes over the specified interval.
Input status indication function Input status indication through the LEDs of each channel function
When in operation:
When overdriven:
Illuminates in green when pulse input is detected
Illuminates in red when the input voltage exceeds the range
Compatible probes/cables Connection cable (1:1): Recommended 1
366926
Voltage probe: Recommended 2
700929 (10:1 safety probe) .20 to 45 pF:
701947 (100:1 probe) .15 to 45 pF:
For measuring 1000 V
(DC+ACpeak) or less
For measuring 1000 V
(DC+ACpeak) or less
1
V
alue measured under standard operating conditions.
H
L
Combined with the 700929 or 701947
2
700929
701947
H
L
Combined with the (701901 + 701954) or
direct input (cable that does not comply with the safety standards)
4
5
BNC
Withstand voltage: 1500 Vrms for 1 minute
Allowable transient surge voltage (between earth and input): ±2100 Vpeak
3
6 Input waveform of 1 Vpp, rectangular wave, rise/fall time within 1 ms (input range: ±10 V, bandwidth limit: Full, and hysteresis:
±1%)
7 Input waveform of 90 Vrms, sine wave (input range: AC100 V, bandwidth limit 100 kHz, and hysteresis: ±1%)
8 Input waveform of 1 Vpp, rectangular wave, rise/fall time within 5 ns (input range: ±10 V, bandwidth limit: Full, and hysteresis:
±1%)
9 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
Do not apply input voltage exceeding the maximum input voltage or allowable common
mode input voltage.
T
o prevent the possibility of electric shock, be sure to furnish protective earth grounding of
the DL850/DL850V.
T
o prevent the possibility of electric shock, be sure to fasten the module screws.
Otherwise, the electrical and mechanical protection functions will not be activated.
Avoid continuous connection under an environment in which the surge voltage may occur.
6.13 Module Specifications