6-41
IM DL850-03EN
Specifications
High-Speed 100 MS/s, 12-Bit Isolation Module (720210) Specifications
Item Specifications
Standard operating conditions Temperature:
Humidity:
23 ±5°C
20 to 80% RH
After a 30-minute warm-up and after calibration
Effective measurement range 20 div (display range: 10 div)
Number of input channels 2
Input coupling AC, DC, and GND
Maximum sample rate 100 MS/s
Input format Isolated unbalanced
Frequency characteristics
1
(−3 dB point when sine wave of amplitude ±3 div is applied) DC to 20 MHz
Voltage-axis sensitivity setting 10 mV/div to 20 V/div (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or less)
Combined with the 700929 (10:1) or 701947 (100:1):
2
1000 V (DC+ACpeak) CAT II
Direct input (cable that does not comply with the safety standards):
4
200 V (DC+ACpeak)
Maximum allowable common
mode voltage
(at a frequency of 1 kHz or less)
Working voltage of safety standard
Combined with the 700929 (10:1) or 701947 (100:1):
3
1000 Vrms (CAT II)
Direct input (cable that does not comply with the safety standards):
5
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis accuracy
DC accuracy
1
10 mV/div to 20 V/div: ±(0.5% of 10 div)
Input connector BNC connector (isolated type)
Input impedance 1 MΩ ± 1%, approx. 35 pF
−3 dB point when AC coupled
low frequency attenuation point
10 Hz or less (1 Hz or less when using the 700929, 0.1 Hz or less when using the 700929)
Common mode rejection ratio 80 dB (50/60 Hz) or more (Typical
6
)
Residual noise level
(Input section shorted)
±1.1 mV or ±0.15 /div whichever is greater (Typical
6
)
Withstand voltage 1500 Vrms for 1 minute (across each terminal and earth) (60 Hz)
Insulation resistance 500 VDC, 10 MΩ or more (across each input terminal and earth)
A/D conversion resolution 12 bit (150 LSB/div)
Temperature coefficient Zero point:
Gain:
10 mV/div to 20 V/div: ±(0.1% of 10 div)/°C(Typical
6
)
±(0.05% of 10 div)/°C(Typical
6
)
Bandwidth limit Select from OFF, 2 MHz, 1.28 MHz, 640 kHz, 320 kHz, 160 kHz, 80 kHz, , 40 kHZ, 20 kHz,
and 10 kHz
Cut-off characteristics: −12 dB/OCT (when 2 MHz, Typical
6
)
Probe attenuation setting Voltage probe:
Current probe:
1:1, 10:1, 100:1, 1000:1
10 A:1 V (for the 700937/701933), 100 A: 1 V (for the 701930/701931)
1 V
alue measured under standard operating conditions.
H
L
Combined with the 700929 or 701947
2
700929
701947
H
L
Direct input (cable that does not comply with the safety standards)
4
5
BNC
Withstand voltage: 1500 Vrms for 1 minute
Allowable transient surge voltage (between earth and input): ±2100 Vpeak
3
6 Typical value represents a typical or average value. It is not strictly warranted.
6.13 Module Specifications