6-43
IM DL850-03EN
Specifications
1
2
3
4
5
6
App
Index
16-CH Voltage Input Module (720220) Specifications
Item Specifications
Standard operating conditions Ambient temperature: 23 ± 5°C
Ambient humidity:
20 to 80% RH
After the DL850/DL850V has been warmed up for 30 minutes and then calibration
has been performed
Valid measurement range 20 div (0 V ± 10 div. Display range: 10 div. When Variable is set to OFF)
Number of input channels 16
Input coupling settings DC, GND, and OFF
Maximum sample rate 200 kS/s for a single channel, 16 ch × 10 kS/s
Input format All isolated unbalanced
Frequency characteristics
1
DC to 5 kHz
(−3 dB point when a sine wave with an
amplitude of ±3 div is applied)
Selectable voltage sensitivity range 0.2 V/div to 2 V/div (in 1-2-5 steps)
Max. measurement voltage range ±20 V
Maximum input voltage
2
Direct input 42 V (DC + ACpeak)
(at a frequency of 1 kHz or less)
Maximum allowable common mode voltage
3
The Working Voltage in the safety standards
(at a frequency of 1 kHz or less) Direct input
42 V (DC + ACpeak) (CAT II, 30 Vrms)
DC vertical-axis (voltage-axis) accuracy
1
± (0.3% of 10 div)
Input connector Spring terminal blocks. Each 8-channel unit is removable.
Input impedance 1 MΩ ± 1%
Common mode rejection ratio 80 dB (50/60 Hz) or more (typical value
4
)
Residual noise level (input section shorted) ±0.05 div
A/D converter resolution 16 bits, 2400 LSB/div
Temperature coefficient Zero point: ±(0.02% of 10 div)/°C (typical value
4
)
Gain: ±(0.02% of 10 div)/°C (typical value
4
)
Bandwidth limit Can be turned off or set to 500 Hz on each sub channel
Wiring Recommended: 0.20 mm
2
to 1.00 mm
2
. AWG size: 24-18.
1 Value measured under standard operating conditions.
H
L
3.
2.
4 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
When connecting a device under measurement to the instrument, be sure to turn off the
device. It is extremely dangerous to connect or remove wires while the device under
measurement is on.
Avoid Electric Shock When Using the Modules
Do not apply a voltage that exceeds the maximum input voltag
e, withstand voltage, or
allowable surge voltage.
T
o avoid electric shock, be sure to ground the instrument.
To prevent the possibility of electric shock, be sure to fasten the module screws. Otherwise,
the electrical and mechanical protection functions will not be activated.
Do not leave the instrument connected to devices in an enviro
nment that may be subject to
voltage surges.
T
o prevent electric shock, connect wires to the terminal block that match the voltage range
that you are measuring.