CAUTION
Correct measurements cannot be obtained when the positive and negative thermocouple
leads are connected in reverse.
Immediately after connecting the thermocouple, the heat balan
ce may be disturbed at
the input terminal section and may cause measurement errors. Therefore, wait about ten
minutes before making a measurement.
In an environment where the air from the air conditioning is dir
ectly applied to the input
terminals or where there are effects from a heat source, the heat balance may be disturbed
at the input terminal section and cause measurement errors.
When making measurements in this type of environment, take preventive measures such
as changing the position.