App-29
IM DL850-03EN
Appendix
1
2
3
4
5
6
App
Index
Item Specifications
Compatible probes/
cables
Connection cable (for high voltage 1:1): Recommended 1
701901 (isolated type BNC-safety alligator clip adapter ×2: For measuring 850 V (DC+ACpeak) or
less), 701954 (alligator clip (dolphin type) red/black 2-piece set) is required separately
Voltage probe: Recommended 2
700929 (10:1 safety probe).20 to 45 pF: For measuring 1000 Vpeak or less
701947 (100:1 probe).15 to 45 pF: For measuring 1000 Vpeak or less
Current probe (power can be supplied from the DL850/DL850V. Option)
700937 (15 A), 701930 (150 A), 701931 (500 A), 701933 (30 A)
1 V
alue measured under standard operating conditions.
H
L
Combined with the 700929 or 701947
2
3
4
700929
701947
H
L
701901
701954
5
6
7
Combined with the 701901+701954
H
L
Direct input (cable that does not comply with the safety standards)
8
9
BNC
Withstand voltage: 3700 Vrms for 1 minute
Allowable transient surge voltage: ±5200 Vpeak
(between earth and input)
10 The typical value is a representative or standard value. It is not strictly warranted.
WARNING
When applying high voltage using this module, use the 1:1 safety cable (combination of
701901 and 701954) or the isolated probe (700929 or 701947).
Note that the overvoltage tolerance of the low and high sides o
f this module may differ
depending on the combination of the connected probe, measurement lead, and clip.
Do not apply input voltage exceeding the maximum input volta
ge, withstand voltage, or
allowable surge voltage.
T
o prevent the possibility of electric shock, be sure to furnish protective earth grounding of
the DL850/DL850V.
T
o prevent the possibility of electric shock, be sure to fasten the module screws. Otherwise,
the electrical and mechanical protection functions will not be activated.
A
void continuous connection under an environment in which the surge voltage may occur.
Appendix 6 High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701260) Specifications