Phase Compensation Signal
The instrument generates the following square wave signal from the COMP signal output terminal.
Frequency:
1 kHz ± 1%
Computation: 1 V ± 10 %
Differences in the Waveform due to the Phase Correction of the
Probe
Correct waveform
Too much adjustment
(increased gain in the high
frequency band)
Not enough adjustment
(reduced gain in the high
frequency band)
Notes about Using Probes with the Frequency Module (701280
(FREQ))
You cannot perform phase correction on a probe connected to the frequency module. To use a probe
with the frequency module, first perform phase correction on the probe using another module.
3.6 Correcting the Probe Phase
3-21
IM DL850-03EN
Making Preparations for Measurements
1
2
3
4
5
6
App
Index

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL850 ScopeCorder and is the answer not in the manual?

YOKOGAWA DL850 ScopeCorder Specifications

General IconGeneral
Maximum Sampling Rate100 MS/s
Power Supply100 to 240 VAC, 50/60 Hz
Resolution16-bit
Operating Temperature5 to 40°C
Bandwidth20 MHz
Input TypesVoltage, Temperature, Strain, Acceleration, Frequency, Logic, Current
InterfacesEthernet, USB, GP-IB
Trigger ModesEdge, Window, Pattern
Number of ChannelsUp to 128 (with multiple input modules)
Memory CapacityUp to 2 G points (with /M4 option)

Related product manuals