5 First Operating Steps | 5.9 De-energizing the Microscope ZEISS
5.9 De-energizing the Microscope
This procedure completely cuts off the microscope from the electrical main supply.
WARNING
Residual voltage at the mains plug
After unplugging the mains plug residual voltage is present at the pins of the plug which may
cause electrical shock.
4 After unplugging the mains plug wait at least 5 s before touching the pins of the mains
plug.
Prerequisite
ü
The microscope is in Standby mode or in OFF mode, refer to Finishing the Work Session | LaB₆
Filament [
}
108] or Finishing the Work Session | Tungsten Filament [
}
108].
Procedure
1. If the microscope is not yet in OFF mode, press the Off button at the front of the plinth.
à
When all subsystems are fully deactivated,
the Off button lights up red permanently.
2. Close and lock the main shut-off valves at the installation site.
3. If the EMO circuit is not installed, unplug the
power cord by unplugging the CEE connector
from the CEE FEMALE RECEPTACLE of the
mains supply.
4. If the EMO circuit is installed, switch the Main
Switch to its OFF position.
9 The microscope is completely cut off from the electrical main supply.
110 Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006