EasyManua.ls Logo

Zeiss EVO - Setting up the BSD Detector

Zeiss EVO
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ZEISS 5 First Operating Steps | 5.6 Finding Appropriate Detector Settings
Info
The C2DX detector requires VP pressures of at least 20Pa to acquire a good signal.
A good starting point is EHT=20kV, I Probe=500pA, Chamber pressure=30Pa,
WD=8.5mm, and scan speed=5.
Procedure
1. In the SEM Controls panel, select the Detectors tab.
2. From the Signal A drop-down list, select C2DX.
3. In the Panel Configuration Bar, select C2DX Control.
à
The C2DX Control panel is displayed.
4. Select C2DX Gain = Low.
5. Adjust the C2DX Bias scroll bar until you obtain a good and stable image.
INFO: A typical value for C2DX Bias is 50%. Contrast and brightness should also be ad-
justed to obtain an image.
6. Click C2DX Auto Level.
7. If you are unable to reach saturation or if the image quality is not good, select C2DX Gain
= High and repeat the procedure.
5.6.5 Setting up the BSD Detector
Prerequisite
ü
Chamber door is open.
Procedure
1. Check that the lens-mounted BSD detector is fitted.
2. If it is in the parked position, remove the BSD
detector from the parked position.
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006 85

Table of Contents

Related product manuals