MODULE 4: POWER AND CIRCUIT BOARDS
4-6 ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY 201532-102
The RSH contains jumpers, fuses and test points. The RSH Card Cage is jumpered at JU1 and JU2 for
recognition of the RSH configuration.
A2_CBT059a
J41
J48
J678J679
J71
J70
J681J680
J42
1
5
9
6
F2
F1
10A
2A
TP1 +5VDC
TP2 +36V
TP3 +5VDC
J600
J602
J46
J701
J721
J722
J702
J616
J703
J603
36
1
33
68
J60
36
1
33
68
TP2 GND
TP3 5VDC
TP1 36VDC
F1 10A 5V Fuse
F2 2A 12V Fuse
JU2
JU1
RSH Card Cage