MODULE 2: BASIC OPERATION
ARCHITECT
®
RSH Service Training Guide 2-23
201532-102 FOR INTERNAL USE ONLY
Bar Code Troubleshooting
If the RSH Bar Code Reader is not recognizing an empty Sample Carrier Position or empty carrier
positions are detected as bad bar code labels, consider the following:
• Perform the Set Bar Code Transitions procedure (P-226) to edit the expected bar code read
transition field of the
file SQBCR.CAL or LQBCR.CAL to a value correct for the bar code
type and number of characters being used.