MODULE 2: BASIC OPERATION
ARCHITECT
®
RSH Service Training Guide 2-9
201532-102 FOR INTERNAL USE ONLY
BASIC OPERATION
Sample Volume Requirements
The sample volume requirement for routine testing of controls and patient samples is
150µL (minimum)
This recommendation supports onboard sample storage for three hours under average laboratory
conditions without observable concentration effects due to sample evaporation. Reliability of assay results
cannot be guaranteed if this recommendation is not followed. The sample volume required for testing can
be located on the ORDERLIST Report.
NOTE: For additional information on sample volume, refer to the ARCHITECT
System Operations Manual.
Sample Vessel Requirements
The following sample vessels (cup or tube containing sample) are acceptable for use on the ARCHITECT
System:
ARCHITECT System sample cups
ARCHITECT System sample cups used in conjunction with a sample tubes
Aliquot and primary sample tubes
Sample Cup
Sample Cup in bar code-labeled tube
Bar Code
labeled tubes
Sample Carrier
125 µL
500 µL
1400 µL
Carrier
Bar Code ID
Front View
Sample Gauge Label