MODULE 5: MISCELLANEOUS
5-6 ARCHITECT
®
RSH Service Training Guide
FOR INTERNAL USE ONLY 201532-102
INSTALLATION
ARCHITECT RSH Installation procedures are designed to install RSH system hardware on to a standalone
ARCHITECT
®
c8000
®
or i2000
SR
®
System and verify the system hardware is performing within
specification.
Use the ARCHITECT RSH and corresponding ARCHITECT System Service and Support
Manual, Section 7 Pre-Site Interview and Inspection and Section 8 Installation to view the
checklists discussed below.
General Installation
The Installation Checklist can be located in the ARCHITECT RSH System Service and Support Manual,
Section 8 Installation. It contains the procedures that are required to, unpack, setup, and verify
performance. Follow checklist for efficiency and proper sequence.
Installation consists of:
• Unpacking RSH
• Hanging the RSH on the Processing Module
• Installing the RSH Card Cage and corresponding boards (c8000
®
Standalone System only)
• Connecting Cables from Card Cage to RSH
• Installing RSH covers and end caps
• Calibration of robotic positions
• Performance Verification Test
Installation Tips:
• RSH Clamps too loose or not seated in PM pocket properly will result in Sample Pipettor
Calibration failure; the probe misses the target.
• Ensure the RSH is aligned left to right of PM prior to seating and tightening clamps.
• Lubricate pulley prior to tightening belt.
• Ensure belt is properly placed in bearing (full back position at the black timing mark) and
pulley is fulled extended for proper belt tension.