MODULE 3: ROBOTICS
ARCHITECT
®
RSH System Service Training Guide 3-11
201532-102 FOR INTERNAL USE ONLY
Diagnostic Information:
• Carrier Transport must be removed and installed at the extreme left end behind Bar Code
Reader to avoid damaging Section Align Sensor
• M&D 1119 Transport Calibration Procedure aligns the Carrier Transport position with
positioner pockets, tray and Carrier positions
• Carrier Transport Belt must contain a minimum of 10 evenly spaced cords
• Bearing felt wiper pads are replaceable and should be super saturated with SuperLube oil
• Proper Belt positioning is important for proper operation
• J702 and J701 communication cables are identical cables and can be mispositioned during
installation to create errors
• Carrier Transport Flex Board delivers +5VDC to the Rail Guide Sensor Board
• Motors receive 36VDC from the RSH Card Cage
• Rail Guide Sensor Board is replaceable