7351 Series Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-25
Item Command Description
Initial
value
R6451
Calculation
NULL calculation NL0 OFF
NL1 ON
NL? Reply: NL0 or NL1
-
KNLn NULL constant setting
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
Note:
The setting is disabled when the NULL calcula-
tion is set to OFF.
(0)
KNL? Reply: KNL
d.dddddEdd *2
-
Smoothing calcula-
tion
SM0 OFF
SM1 ON
SM? Reply: SM0 or SM1
-
TIn Smoothing count
n: 2 to 100 (counts)
(10)
TI? Reply: TIddd
-
Scaling calculation SC0 OFF
SC1 ON
SC? Reply: SC0 or SC1
-
KAn Constant A (Zero cannot be set).
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(1)
KBn Constant B
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(0)
-
KCn Constant C
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(1)
-
KAM Sets Constant A to the measurement value.
KBM Sets Constant B to the measurement value.
KCM Sets Constant C to the measurement value.
KA? Reply: KAd.dddddEdd *2
-
KB? Reply: KBd.dddddEdd
-
KC? Reply: KCd.dddddEdd
-
*2: The position of the decimal point in the reply is fixed.