7351 Series Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-26
Calculation dB/dBm calculation DB0 dB calculation OFF
DB1 dB calculation ON
For the voltage and current measurement functions, ON
can be set.
DB2 dBm calculation ON
For the voltage measurement function, ON can be set.
DB? Reply: One of DB0 to DB2
-
KDn Constant D
n:0.000001E-9 to 999999.E+6
(1)
KDM Sets Constant D to the measurement value.
KD? Reply: KDd.dddddEdd *2
MAX•MIN
calculation
:
Header (Refer to
Section 6.7.2,
“Output Data
Format.”)
MN0 MAX•MIN calculation OFF
MN1 MAX•MIN calculation ON
*3
MN? Reply: One of MN0 to MN1
-
MAX? MAX value reading *2
Reply: M d.dddddEdd
-
MIN? MIN value reading *2
Reply: I d.dddddEdd
-
AV E ? AV E v a l u e r e a d i n g * 2
Reply: A d.dddddEdd
-
AVN? Measurement count reading *2
Reply: AVNd.dddddEdd
-
Comparator
calculation
CO0 OFF
CO1 ON
CO? Reply: CO0 or CO1
-
HIn HIGH constant
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(0)
LOn LOW constant
n:-999999.E+6 to +999999.E+6
Setting resolution:0.000001E-9
(0)
HIM Sets the HIGH constant to the measurement value.
LOM Sets the LOW constant to the measurement value.
HI? Reply: HId.dddddEdd *2
-
LO? Reply: LOd.dddddEdd *2
-
*2: The position of the decimal point in the reply is fixed.
*3: MN1/MN2 (MAX calculation ON/MIN calculation ON)
Item Command Description
Initial
value
R6451