7.3.1 PSV test performing
1. Connect air circuit
a. Connect the device to the air circuit;
b. Press the Start icon on the right of the screen to start performing test, apply the pressure gradually to the safety
valve;
c. Count down the test time and record the real-time pressure value and create the pressure curve, then record the
maximum pressure during the process;
d. After the countdown completes and the entire leak test process finishes, save the data;
2. The whole process will be shown in the leak test curve at the bottom of the screen
7.4 Thermal calculator
The calibrator provides thermal calculator app, dedicated to variety of industrial TC, RTD and customized types of
sensors. Using the thermal calculator, it is easy to convert between temperature and electrical signals of different
sensors.
7.5 Sensor library
In order to meet the needs of custom sensor types, the calibrator provides a sensor library function. In the sensor library
function, users can define new sensor types according to their needs and set relevant parameters in the sensor. The
sensor library supports a total of four types of custom sensors: ITS-90, CVD, RTD and TC.