1.2.4 Simulate RTD measure & output accuracy (Environmental temp.: 20±10°C)
Table 5 Simulate RTD measure & output
Sensor type Temperature(°C) Annual accuracy(°C)
PT10(385)
-200850
-200200
0.59
200600
0.72
600850
0.82
PT25(385)
-200850
-200200
0.27
200600
0.35
600850
0.41
PT50(385) -200850
-200200
0.16
200600 0.22
600850 0.27
PT100(385)
PT100(391)
PT100(3916)
PT100(3926)
-200850
-200200 0.10
200600 0.16
600850 0.20
PT200(385) -200850
-200200 0.32
200300 0.34
300600 0.41
600850 0.48