Click the "Settings" menu item in the menu of the simulate RTD output to enter the simulate RTD setting menu. The
setting menu provides a variety of parameters for the function.
Table 17 Simulate RTD Settings
Subject Valid Value Description
Sensor Type RTD sensor Select RTD sensor
Temp Unit K / °F / °C Select the temperature unit as Kelvin, Fahrenheit or Celsius
Decimal digits 0~0.001 Decimal accuracy of temperature display
3.5.7 Set point output
Click the signal output value in channel 2 of the calibrator and input the set point value you want to output in the pop-
up set point input keyboard and click the confirm button to set the output value.
3.5.8 Automatic ramp output
The three electrical signal outputs (voltage, current, and frequency of the calibrator) and the simulate RTD/ TC output
all support automatic ramp output mode. The Ramp can increase or decrease the output value smoothly and
continuously. In fact, a small step value is used to form a Ramp signal, which is represented by a Ramp gradient in the
calibrator. The calibrator will automatically calculate the Ramp gradient according to the setting start and end values of
the Ramp and the rise and fall time. When the start value and end value are fixed, the longer the ramp time, the slower
the ramp output and the smaller the ramp gradient.
A complete ramp output stroke includes an up stroke (from the start value ramp to the end value) and a down stroke