63
Click the step range value in the middle of the left and right buttons under the channel to pop up the range input
interface, where the user can customize the output range of the manual step; the customized step output range cannot
exceed the output range of the current signal option.
The step length setting menu button of manual step is at the bottom right of channel 2, click this button to select the
step length of 25%, 100%, or manually input a custom step length by the user.
Click the button at the top right of channel 2 to exit manual step mode.
3.5.11 Fine adjustment output
The electrical signal outputs (voltage, current, and frequency of the calibrator) and the simulate RTD/ TC output can all
be output in the fine-adjustment output mode. Open the setting menu of the corresponding signal and click the "fine
adjustment" menu item to enter the fine adjustment output mode. In the fine-adjustment output mode, users can adjust
the output value incrementally as needed.
After entering the fine-adjustment mode, click the left and right buttons below the channel to move the cursor
on the output value, and click the up and down buttons on the right side of the channel to adjust the value of the
current cursor position. In the process of adjusting the output value using the fine-tuning output mode, the output value
cannot exceed the output range of the current signal option.
Click the button on the upper right of channel 2 to exit the fine-adjustment output mode.