6. Task
The calibrator provides a powerful task function, which makes fully automatic calibration possible. At the same time, the
calibration data can be automatically collected, stored, and analyzed, and it can be easily executed and viewed
repeatedly. Click the task icon on the main interface of the device or click the task button at the top right in the
calibrator interface to enter the task interface.
The documenting task provides task types for four types of equipment: pressure, loop circuit, temperature, and flow.
There are multiple detailed classifications under each task type, and a total of 16 types of DUTs are supported. The
process of using the document task function is as follows:
(1) Create task: Click the add task button at the top right of the interface, and click the task type you want to create in
the pop-up inspection type selection interface.
(2) According to the task creation interface presented on the interface, input the information needed to create the task
accurately and in detail (for example: task name, inspected range, etc.). After completing the information, click the OK
button to save the task information. The tasks created after the task information is saved will be added to the task list
in the main interface of the document task.
(3) Select the task just created and click the "Execute" button in the pop-up interface to run the task. Complete the
configuration required for task operation and execute the task according to the guidance of the task operation interface.
General tasks can choose to run automatically or manually.