Figure 25 Automatic Step Output Settings
1. Stroke: The stroke mode in the automatic step output mode is divided into one-way mode and round-trip mode. In the
single-pass mode, the calibrator outputs signal in sequence from the first output point to the last output point as a
complete stroke; in the round-trip mode, the calibrator outputs signal in sequence from the first output point to the last
output point, and then from the last output point to the first output point, output in turn, and the two together form a
complete stroke.