100
Refer to Finish the task in paragraph 6.1.1.
6.4.3 Signal isolator/ converters
1. Connect to DUT
Connect the DUT input with the output module of the calibrator, as shown in the figure 22 (for current output) or 21
(for voltage output). Connect the DUT output with the electrical measurement module of the calibrator according
to the DUT signal type, as shown in figure 8(Current Measurement) and figure 7(Voltage Measurement).
2. Create the task
Create the corresponding DUT tasks as described above and fulfill the necessary information. See table below.
Table 36 Signal isolator task setting
Subject Valid value Description
Name Numbers, letters, symbols Task name
S/N Numbers, letters, symbols Task S/N
Model Numbers, letters, symbols Model of DUT
Input (4~20)mA,(0~10)mA,(0~20)mA,(1~5)V,(0~5)V,(0~10)V, custom Input range of DUT
Output (4~20)mA,(0~10)mA,(0~20)mA,(1~5)V,(0~5)V,(0~10)V, custom Output range of DUT
Accuracy 0.05%,0.1%,0.2%,0.5%,1%,1.5%,2%,2.5%,custom Accuracy of DUT
Transfer function Linear / square / square root Transfer function of the DUT
Owner Numbers, letters, symbols Owner of DUT