Advanced Automated Osmometer User’s Guide
98
• Select Yes to delete all of the data that is presently being stored.
A message will confirm that the data was deleted, and the A
2
O
will be restarted.
Shutdown and storage
CAUTION Purge the pipettor system and remove all other liquids
from inside the instrument if the ambient temperature will
be below 0ºC (32ºF) at any time during the shutdown or
storage.
Overnight or weekend:
1. Prime the pipettor system.
2. Remove all primary sample tubes, used osmometer sample cups,
and used wiper rings.
3. Empty the waste bottle.
4. Wipe up any spills on the work surfaces.
5. Close the front door.
6. Turn off the instrument, if desired.
CAUTION Do not cover the instrument unless it is turned off.
Instrument overheating or damage to the cooling fans
can result.