4. The instrument checks that the calibrators are in the correct posi-
tions.
5. When the calibrators are correctly identified, the instrument automat-
ically runs five replicates from each primary tube.
6. After the calibration routines are complete, the updated calibration
status Calibration OK is displayed. If, for any reason, the calibra-
tion is not successfully completed or is exited prior to completion,
the prior calibration status screen will be displayed.
7. To verify the Low Range calibration, run samples of Clinitrol™
Reference Solution (290). To verify the Full Range, run samples of
1500 mOsm Calibration Standard.
Calibration notes
The A
2
O retains its previous calibration data until it completes a
recalibration and displays Calibration Completed.
It is your responsibility to place the required calibration standards in
the correct order into the primary carousel.
43
Calibration
Figure 11: The Calibrate Screen