EasyManua.ls Logo

Advanced Instruments A2O - Page 66

Advanced Instruments A2O
157 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model A
2
O
®
Advanced Automated Osmometer Users Guide
44
Pressing Stop during calibration cancels the current calibration test.
Pressing Stop a second time will cancel calibration and retain the
previous calibration.
Only calibration testing can be performed when a calibration is in
progress. Testing of reference materials, controls, or unknown sam-
ples must be performed using the normal test procedure (see Chapter
2).
The instrument must successfully test most replicates at each stan-
dard level. The instrument will still calibrate if one or two errors
occur. If the repeatability is acceptable on these samples, the instru-
ment will automatically perform the internal calibration. Successful
calibration is indicated by the message “Calibration completed”.
NOTE If you have changed the block probe or sample probe resis-
tances, you will see the message Recalibration Needed”. Pay
close attention to the displayed prompts. You will be requested
to run a slightly different sequence than described above.

Table of Contents

Related product manuals