EasyManuals Logo

Advanced Instruments A2O User Manual

Advanced Instruments A2O
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
101
Chapter Name
Troubleshooting Table
Appendix A
Problem/Message
Explanation
Abrupt loss of power
“Block Probe Open?”
“Calibration out of
Range”
“Cannot Read Barcode”
If you lose power to your instrument, we rec-
ommend that you check that your outlet is
providing the correct amount of power.
Check that your cord is firmly plugged into
both the instrument and the outlet. Visually
inspect the fuses, and change as necessary.
If you see this message, first try restarting
your instrument. Contact Advanced
Instruments Hot-Line Service for instructions
on verifying that the block probe is plugged
into the osmometer circuit board.
If you see this message, verify that there is
sufficient fluid in the calibration ampules.
Then recalibrate. If the problem persists,
contact Advanced Instruments Hot-Line
Service.
Primary sample tubes or calibration adapters
do not contain a barcode, or they are posi-
tioned so that the barcode cannot be read.
Manually enter the barcodes, or reposition
the tubes and holders so that the barcodes are
visible in the slots of the tube holders.
Make sure that the instrument is unplugged before remov-
ing the instrument cover!
NOTE Messages below may appear in message screens or error logs.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments A2O and is the answer not in the manual?

Advanced Instruments A2O Specifications

General IconGeneral
BrandAdvanced Instruments
ModelA2O
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals