EasyManuals Logo

Advanced Instruments A2O User Manual

Advanced Instruments A2O
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
1
Installation & Setup
To set up your instrument properly, it is important that you read
and follow the steps in this section. Please follow these steps care-
fully and be sure to read Chapter 2 — Instrument Operation
before attempting to run tests on your instrument. For those per-
forming Installation Qualification (IQ) and Operation Qualifica-
tion (OQ), please contact Advanced Instruments or authorized
dealer for more detailed documentation available.
Step 1 — Find a location for the instrument
When choosing a location for your new osmometer, be sure to
meet the following criteria:
Adequate space. The dimensions of the instrument are 23.6 ×
22.8 × 20.5 inches (60 × 58 × 52 cm). Be sure to keep your
workplace free of debris and allow 6 inches (15 cm) of clear
space on both sides and behind the instrument for fan-driven
air circulation.
Electric outlet availability. Your instrument will need to oper-
ate within five feet of a properly grounded, three-prong elec-
trical outlet capable of continuously supplying 3 amperes at
100-250V. If the instrument is not grounded properly, its
operation may be impaired and it may be a safety hazard.
Therefore, be sure to test the outlet and record the results
before operating your instrument.
1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments A2O and is the answer not in the manual?

Advanced Instruments A2O Specifications

General IconGeneral
BrandAdvanced Instruments
ModelA2O
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals