Model A
2
O
®
Advanced Automated Osmometer User’s Guide
108
Performance Over
Operating Conditions
Temperature Recalibration recommended for ambient temperature changes
Effect: of 5°C (9°F) or greater
Operating
Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Humidity: 5 to 80% relative humidity (non-condensing)
Sample Viscosity: Up to 20 mPa
.
s (20 cP); higher viscosities may affect
pipettor accuracy
Storage Temp.: -40°C to +45°C (-40°F to +113°F)
Start-up Time: Immediately from stand-by; under 1 minute from power-on
Test Time: Approximately 3 minutes per sample in low range
Sample Throughput: Can process 20 samples in less than 1 hour
Pipettor Accuracy: < 1%
Sound Level: Measured at operator’s position < 66 dB(A) Typ.
Dimensions inches centimeters
Width: 23.6 60
Depth: 20.5 52
Height: 22.8 58
Weight pounds kilograms
Net: 68 31
Shipping: 133 60
Warranty: One-year limited warranty on workmanship and all parts
except glass, plastic, and parts warranted by their makers
Certification:
Refer to Regulatory Notices (Appendix C) for applicable
standards
Installation Class: I
Over-Voltage
Category: II
Pollution Degree: 2
Moisture Protection: IPX0 (ordinary)