NOTES
The plastic sample tube is designed for single use only and must
be discarded after one test. The mechanical impact used to nucle-
ate the sample state change during the test deforms and cracks the
plastic tube. Reusing the cracked tube can result in sample leak-
ing into the cooling chamb er. This contaminates the cooling
chamber. It can also produce inaccurate test results or errors.
Advanced Instruments cannot guarantee the stated instrument per-
formance specifications unless Advanced Instruments brand con-
sumables are used with the instrument. Use of consumables from
manufacturers other than Advanced Instruments is not recom-
mended and may adversely affect system calibration, perfor-
mance, and operation. For information on obtaining these stan-
dards and controls, contact Advanced Instruments or an autho-
rized representative.