Chapter 4 Maintenance Procedures92
USG-CUS-059 Rev05MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
3. Tap SONICATOR SOLUTION Fill (Figure 4.3).
The pin array head moves to allow access to the ultrasonic wash.
4. Open the main door.
The safety interlock is disengaged! message appears.
5. Manually fill the ultrasonic wash station with 100% ethanol (Figure 4.5).
Figure 4.2 Draining the Ultrasonic Wash
Figure 4.3 Maintenance Screen
Figure 4.4 Safety Interlock Disengaged Message