1290 Infinity Quaternary Pump User Manual 109
Error Information
7
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable cause Suggested actions
1
Leak detected in another module with a
CAN connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Leak detected in an external instrument
with a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
3
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down
condition.
4
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals