198 1290 Infinity Quaternary Pump User Manual
9
Maintenance
Replacing the Outlet Filter
Replacing the Outlet Filter
When For removing blockages and leaks in the high pressure filter assembly. The outlet filter should be
replaced as required depending on the system usage.
Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-1924 Open-end wrench 14 mm
Tor q u e w r e nch
Torque wrench head, 14 mm for torque wrench
Parts required p/n Description
G4204-60004 Outlet filter Quaternary Pump/Flexible Pump
1
Remove the capillary from the outlet filter to the
pressure sensor.
2
Remove the outlet filter using a 14 mm wrench.