1290 Infinity Quaternary Pump User Manual 119
Error Information
7
Pump Error Messages
Reading the pump encoder tag failed
Error ID: 29201
Reading the pump encoder tag has failed.
• Parameter: 1 – 2 referring to pump drive
Writing the pump encoder tag failed
Error ID: 29200
Writing the pump encoder tag has failed.
• Parameter: 1 – 2 referring to pump drive
Pump drive blocked or encoder failed
Error ID: 29214
Pump drive blocked or encoder failed.
• Parameter: None
Probable cause Suggested actions
1
Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2
Missing or defect tag Defect connection
between tag and encoder.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2
Defect tag Defect connection between tag
and encoder.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Blockage of the pump drive Drive encoder
failed.
Please contact your Agilent service
representative.