1290 Infinity Quaternary Pump User Manual 59
Using the Module
4
Best Practices
How to deal with solvents
Use clean bottles only.
Exchange water-based solvents daily.
Select solvent volume to be used up within 1 – 2 days.
Use only HPLC-grade solvents and water filtered through 0.2 m filters.
Label bottles correctly with bottle content, and filling date / expiry date.
Use solvent inlet filters.
Reduce risk of algae growth: use brown bottles for aqueous solvents, avoid
direct sunlight.
Select channels for Multi-Channel Gradient Valve (MCGV)
Use lower channels (A and/or D) for buffer solutions.
Regularly flush all MCGV channels with water to remove possible salt
deposits.
Check compatibility of buffers and organic solvents to avoid precipitation.
CAUTION
Contaminated seal wash solvent
Do not recycle seal wash solvent to avoid contamination.
Weekly exchange seal wash solvent.