122 1290 Infinity Quaternary Pump User Manual
7
Error Information
Pump Error Messages
Deliver underrun
Error ID: 29233
Internal error.
• Parameter: None
Defect connection between main board and pump drive encoder
Error ID: 29208
Defect connection between main board and pump drive encoder.
• Parameter: 1 – 2 referring to pump drive
Pump drive encoder defect
Error ID: 29209
Defect pump drive encoder.
• Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1
Internal error.
Please contact your Agilent service
representative.
2
Firmware issue
Use a minimum firmware revision of B.06.55
Probable cause Suggested actions
1
Defect connection between main board and
pump drive encoder.
Please contact your Agilent service
representative.
2
Defect encoder.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defect encoder.
Please contact your Agilent service
representative.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals