1290 Infinity Quaternary Pump User Manual 117
Error Information
7
Pump Error Messages
Target pressure not reached for quaternary pump degasser
Error ID: 29221
The target pressure of the quaternary pump degasser has not been reached
within the expected time.
Parameter: Pressure in mbar
Solvent counter exceeded limit
Error ID: 29146
The counter for the solvent volume has exceeded the limit, which has been set
in the user interface.
Waste counter limit exceeded
Error ID: 29147
The counter for the waste volume has exceeded the limit, which has been set
in the user interface.
Parameter: None
Probable cause Suggested actions
1
Condensation in degasser chamber due to
temperature fluctuation.
Equilibrate and restart module.
2
Degasser is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
No solvent present.
Refill solvent bottle.
2
Inappropriate setting.
Check solvent counter setting in user interface.
Probable cause Suggested actions
1
The waste container is full.
Empty waste container.
2
Inappropriate setting for waste counter.
Reset waste counter.
Adjust waste counter limit.