EasyManuals Logo

Agilent Technologies 1290 infinity User Manual

Agilent Technologies 1290 infinity
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
1290 Infinity Quaternary Pump User Manual 117
Error Information
7
Pump Error Messages
Target pressure not reached for quaternary pump degasser
Error ID: 29221
The target pressure of the quaternary pump degasser has not been reached
within the expected time.
Parameter: Pressure in mbar
Solvent counter exceeded limit
Error ID: 29146
The counter for the solvent volume has exceeded the limit, which has been set
in the user interface.
Waste counter limit exceeded
Error ID: 29147
The counter for the waste volume has exceeded the limit, which has been set
in the user interface.
Parameter: None
Probable cause Suggested actions
1
Condensation in degasser chamber due to
temperature fluctuation.
Equilibrate and restart module.
2
Degasser is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
No solvent present.
Refill solvent bottle.
2
Inappropriate setting.
Check solvent counter setting in user interface.
Probable cause Suggested actions
1
The waste container is full.
Empty waste container.
2
Inappropriate setting for waste counter.
Reset waste counter.
Adjust waste counter limit.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals