196 1290 Infinity Quaternary Pump User Manual
9
Maintenance
Replacing Parts of the Multi Purpose Valve
Replacing Parts of the Multi Purpose Valve
1 Use the 9/64 inch hex key for opening the valve head.
2 Replace parts as required.
Tools required p/n Description
8710-2394 9/64 inch hex key
Parts required p/n Description
1534-4045 Bearing ring
5068-0123 Rotor seal, Multi Purpose Valve 1290 Infinity Quaternary Pump, 1200 bar
5068-0120 Stator ring
5068-0001 Stator head
1535-4857 Stator screws, 10/pk
Preparations Remove all capillary connections from the Multi Purpose Valve.