EasyManua.ls Logo

Agilent Technologies 1290 infinity - Best Practices; Daily; Weekly Tasks

Agilent Technologies 1290 infinity
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56 1290 Infinity Quaternary Pump User Manual
4
Using the Module
Best Practices
Best Practices
Daily / Weekly tasks
Daily tasks
Replace solvents and solvent bottles for mobile phases based on
water/buffer.
Replace solvents and solvent bottles for organic mobile phase latest every
second day.
Check presence of seal wash solvent.
Purge each channel with fresh solvent at 2.5 – 3 mL/min for 5 min.
Equilibrate your system with composition of your application for 15 min.
Use conditioning for 1290 Infinity II Pumps and G7104C.
Weekly tasks
Change seal wash solvent (10 % / 90 % isopropanol/water) and bottle.
Flush all channels with water at 2.5 – 3 mL/min for 5 min to remove salt
deposits if buffer applications were used.
Inspect solvent filters for dirt or blockages. Clean or exchange if no flow is
coming out of the solvent line when removed from the degasser inlet.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Related product manuals