1290 Infinity Quaternary Pump User Manual 143
Maintenance
9
Introduction to Maintenance
Introduction to Maintenance
Figure 20 on page 143 shows the main user-accessible assemblies of the
Agilent 1290 Infinity Quaternary Pump. These parts can be accessed from the
front (simple repairs) and don’t require to remove the pump from the system
stack.
Figure 20 Overview of Maintenance Parts
Recommended Interval for Preventive Maintenance
The recommended interval for preventive maintenance is:
• 100 L (150 L for Long Life Technology) or 1 year (whichever comes first).
This recommendation is valid for LC instruments on which “typical”
applications are running.
A “typical” application can be characterized as follows:
• pressure range 100 – 800 bar,
• flow rates 0.5 – 3.5 mL/min,
• typical solvents used in reversed phase LC.