EasyManuals Logo

Agilent Technologies 1290 infinity User Manual

Agilent Technologies 1290 infinity
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
1290 Infinity Quaternary Pump User Manual 25
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tab le 1 Physical Specifications
Type Specification Comments
Weight 15.2 kg (33.4 lbs)
Dimensions
(height × width × depth)
200 x 345 x 435 mm
(8 x 13.5 x 17 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 220 VA / 180 W / 615 BTU/h Maximum
Ambient operating
temperature
4–55 °C (39–131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 2000 m (6562 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.