1290 Infinity Quaternary Pump User Manual 155
Maintenance
9
Replacing the Outlet Valve
Replacing the Outlet Valve
When If Outlet valve is defective.
Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-2603 Spanner-double open ended 12X14 mm Chrome
5067-5688 Torque wrench 1 – 25 Nm with 14 mm wrench
G4220-20041 Bit Torx 10x25 mm
Parts required p/n Description
G4220-60028 Outlet valve
(primary pump head)
G4220-20020 Internal gold seal for Outlet Valve
Preparations Switch off pump at the main power switch
Remove the front cover
Use an optional solvent shutoff valve or lift up solvent filters inside solvent bottles for avoiding
leakages