60 1290 Infinity Quaternary Pump User Manual
4
Using the Module
Best Practices
Optional Inline Filter
The pump can be equipped with an additional inline filter (Inline Filter
Assembly (5067-5407)) with a nominal filter pore size of 0.3 L.
Advantages of the inline filter:
Very small internal volume
Specified for working at high pressures
Possibility of back-flushing the filter
Using the inline filter is recommended:
to protect the downstream system from blockages,
for solvent combinations that can form precipitation after mixing,
for applications running with buffers.
General hints for effective usage of the inline filter:
filter solvents before usage,
follow best practices,
back-flush the filter weekly,
exchange the filter frit on a regular basis.
See Technote G7167-90130 for further reference.
CAUTION
Damage to the valve
Use the filter flush mode only if the optional inline filter is installed.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals