1290 Infinity Quaternary Pump User Manual 57
Using the Module
4
Best Practices
Power up / Shut-down the pump
Power up the pump
Use new or different mobile phase (as required).
Purge each channel with 2.5 – 3 mL/min for 5 min. Open the manual purge
valve or use the purge command.
Equilibrate your system with composition of your application for 15 min.
Use conditioning for 1290 Infinity II Pumps and G7104C.
Long-term shut-down of the system
Flush system with water to remove buffer.
Remove all samples from the sampler and store according to good
laboratory practice.
Use recommended solvents to store the system.
Power off the system.
Prepare the pump
Purge
Use the Purge function to:
fill the pump,
exchange a solvent,
remove air bubbles in tubes and pump heads.
Condition
Use the Conditioning function:
daily when starting the pump,
to minimize pressure ripple by dissolving air bubbles in pump heads.
NOTE
Condition your complete system with solvents and composition of your application (for
example 50 %/50 % A/B at 0.5 mL/min).