EasyManuals Logo

Agilent Technologies 1290 infinity User Manual

Agilent Technologies 1290 infinity
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
1290 Infinity Quaternary Pump User Manual 131
Error Information
7
Pump Error Messages
Seal wash pump was missing when tried to turn on
Error ID: 29223
The seal wash pump has not been detected (while being configured or detected
before)
Valve hardware overcurrent (MCGV)
Error ID: 29227
Power consumption too high for one of the MCGV valves.
Probable cause Suggested actions
1
Defect cable connection to seal wash
pump.
Check cable connection.
2
Defect seal wash pump motor.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Cable defect.
Replace MCGV.
2
Valve defect
Replace MCGV.
3
Defective main board.
Please contact your Agilent service
representative.

Table of Contents